COORD3 ARES
- Ultra rigid alloy frame for maximum rigidity for scanning applications
- Structure with rapid thermal diffusion for changes in environmental conditions
- Bridge beam extrusion designed according to FEA, which allows an optimal moment of inertia for minimum deflection at high accelerations
- M8 threaded table that fits into the large checkerboard system.
- Rigid air bearings for coefficients of high distance between the bearings
- Pneumatic counterbalance of the Z-axis.
- 0.1-micron free-floating resolution measurement scales incorporating dynamic signal processing.
- Fully digital motion control with fusion of the probe path for optimal performance
- Friction reducers with hysteresis close to zero on all axes
- Passive vibration damping system that isolates external vibrations
- Free access to the measuring area from all sides
- Maximum positioning speed: 517 mm / sec
- Maximum acceleration: 1730 mm / sec 2
Each COORD3 CMM is tested according to UNI EN ISO 10360-2.
ARES 05.05 / 07.05

The performances indicated are valid in the following conditions:
- PH10T / M / PH20-TP20 / TP-200: SF module, stylus Ø 4mm x 10mm
- PH6M / PH10M-SP25M : SM1, stylus Ø 5mm x 50mm
- L = measurement length in mm
- Multi-probe thermal compensation (optional)
- Ambient temperature conditions:
T 1: 18 ÷ 22 ° C; Max gradients: 0.5 ° C / h - 2.0 ° C / 24h - 0.5 ° C / m
T 2 : 16 ÷ 26 ° C; Max gradients: 1.0 ° C / h - 5.0 ° C / 24h - 1.0 ° C / m
[1] Indication error of the CMM on dimension measurements according to the ISO 10360-2 standard
[2] Point taking error according to the ISO 10360-2 standard
[3] Scan point taking error according to the ISO 10360-4 standard for SP25M probes
ARES NT 05.05 / 07.05 / 07.07

The performances indicated are valid in the following conditions:
- PH10T / M / PH20-TP20 / TP-200: SF module, stylus Ø 4mm x 10mm
- PH6M / PH10M-SP25M : SM1, stylus Ø 5mm x 50mm
- L = measurement length in mm
- Multi-probe thermal compensation (optional)
- Ambient temperature conditions:
T 1: 18 ÷ 22 ° C; Max gradients: 0.5 ° K / h - 2.0 ° K / 24h - 0.5 ° K / m
T 2 : 16 ÷ 26 ° C; Max gradients: 1.0 ° K / h - 5.0 ° K / 24h - 1.0 ° K / m
[1] Indication error of the CMM on dimension measurements according to the ISO 10360-2 standard
[2] Point taking error according to the ISO 10360-2 standard
[3] Scan point taking error according to the ISO 10360-4 standard for SP25M probes